beautypg.com

Rs-232 settings, Rs-232 pin detail – ETS-Lindgren FM5004 Field Monitor (Archived) User Manual

Page 59

background image

The receiver asserts two lines: Data Terminal Ready (DTR) and Request To

Send (RTS). These lines are asserted continuously and do not perform a

handshake function; they can simply be connected to the CTS and DCD lines at

the connector. This will allow the receiver to transmit with a minimum of wires

and with no handshaking. See RS-232 Pin Detail on page 59 for pinout

information.

Once the RS-232 interface is established, commands are processed in a manner

similar to the IEEE interface. The command structure is identical, except there is

no EOI line; therefore, all commands must be terminated by a .

When a valid command is received, it is processed and the result is immediately

transmitted back over the RS-232 interface. The CTS line can be used to control

the flow of information from the receiver, as described on page 58. The output

format is identical to that of the IEEE format except there is no EOI line.

RS-232

S

ETTINGS

Word Length:

8 bits

Stop Bits:

1

Baud Rate:

1200–19200 (menu selectable)

Parity:

None

RS-232

P

IN

D

ETAIL

Connector: DB-9S

Mating connector:

DB-9PS

Remote Operation

| 59

Archived 10/19/10