2.0 Maintenance

Before performing any maintenance, follow

the safety information in the ETS-Lindgren

Product Information Bulletin included with

your shipment.

Maintenance of the FM5004 is limited to

external components such as cables or

connectors.

Warranty may be void if the housing is

opened.

If you have any questions concerning

maintenance, contact ETS-Lindgren

Customer Service.

WARRANTY

Maintenance Recommendations

W

EEKLY

Check battery status using front panel switch or remote command.

M

ONTHLY

Clean contaminants from the front panel with a damp cloth. Always

unplug the unit before cleaning.

Disconnect fiber optic cables and verify they are clean. See

Maintenance of Fiber Optics on page 18 for cleaning guidelines.

Inspect cables for kinks, breaks, cracks, and loose connectors.