Error code q: sample/s–1 process err–rate too high, Error code r: receive err data not rec from probe, Error code s: send error probe not ready to send – ETS-Lindgren FM5004 Field Monitor (Archived) User Manual
Page 108: Error code u: temp upper temp over upper limits, Error code w: batt warn at warning level, Error code y: alarm upper 4/3 display alarm, Error code z: alarm lower 4/3 display alarm
108 |
Troubleshooting and Error Handling
E
RROR
C
ODE
Q:
S
AMPLE
/
S
–1
P
ROCESS
E
RR
–R
ATE
T
OO
H
IGH
A probe is connected that does not support the current sample rate of
the FM5004. The sample rate can only be set to the rate at which the
slowest probe will operate.
E
RROR
C
ODE
R:
R
ECEIVE
E
RR
D
ATA
N
OT
R
EC
F
ROM
P
ROBE
The FM5004 sent a command to a probe and no response was
received for one sample cycle.
E
RROR
C
ODE
S:
S
END
E
RROR
P
ROBE
N
OT
R
EADY TO
S
END
The FM5004 was ready to send a command to a probe but the
complete response was not received from the previous command.
E
RROR
C
ODE
U:
T
EMP
U
PPER
T
EMP
O
VER
U
PPER
L
IMITS
A temperature measurement caused an upper temperature alarm to
occur.
E
RROR
C
ODE
W:
B
ATT
W
ARN
A
T
W
ARNING
L
EVEL
The battery voltage of a probe fell below the warning level. This
indicates that the probe will operate for only a short time before the
battery fails. Charge the probe battery as soon as possible.
E
RROR
C
ODE
Y:
A
LARM
U
PPER
4/3
D
ISPLAY
A
LARM
A field measurement caused the 4-probe or 3-axis upper alarm to
occur.
E
RROR
C
ODE
Z:
A
LARM
L
OWER
4/3
D
ISPLAY
A
LARM
A field measurement caused the 4-probe or 3-axis lower alarm to
occur.
Archived 10/19/10