beautypg.com

Last errors, Lcd brightness – ETS-Lindgren FM5004 Field Monitor (Archived) User Manual

Page 81

background image

L

AST

E

RRORS

Description:

Returns the code and a description of the last errors

that have occurred.

Syntax:

LERR?

Parameters:

Error message (string):

• Up to 10 lines.

The message contains the error code followed by

a description of the error.

Example:

To see the last error:

LERR?

The FM5004 will return:

00:06:53 C1=H Probe Fail–probe did not recover

If no errors are logged the FM5004 will return:

’No errors logged.’

LCD

B

RIGHTNESS

Description:

Sets the LCD backlight level to a specific brightness.

Syntax:

LCDB,xx

Parameters:

Brightness level (xx):

• Range 0 to 15.

A zero turns off the LCD backlight and a 15 sets

maximum brightness.

Example:

To set the brightness to 8:

LCDB,8

The FM5004 will return:

LCDB,8

Remote Operation

| 81

Archived 10/19/10