Alarm upper value – ETS-Lindgren FM5004 Field Monitor (Archived) User Manual

Page 69

background image

A

LARM

U

PPER

V

ALUE

Description:

Sets the upper levels for the line 1, line 2,

4-probe/3-axis, and high temperature alarms. Each of

the alarm levels must be set using individual

commands.

For an alarm to operate, that alarm must be enabled.

The data entered can contain up to five numbers and a

decimal point.

See the description for each alarm type for the

permissible range. See Alarm Upper Enable on

page 68 to enable an alarm.

Syntax:

AUPV,m,xxxxxx

Parameters:

Alarm Type (m):

• 0 = high temperature alarm,

range: 32

°

F–211

°

F

• 1 = line 1 alarm,

range: 0—max of probe

• 2 = line 2 alarm,

range: 0—max of probe

• 4 = 4-probe/3-axis alarm,

range: 0—max of lowest range probe

Alarm level (xxxxxx)

• xxxxxx = alarm value: 0–1999.9

Example:

• To set the line 1 upper alarm value to 150.5:

AUPV,1,150.5

• To verify the line 1 upper alarm value:

AUPV,1?

The FM5004 will return:

AUPV,1,150.5

Remote Operation

| 69