beautypg.com

Trigger – ETS-Lindgren FM5004 Field Monitor (Archived) User Manual

Page 104

background image

T

RIGGER

Description:

Causes the command contained in the define device

trigger buffer to be executed.

The trigger command (TRG) is useful when used along

with a define device trigger (DDT) command to

continuously execute a set of commands. Entering a

TRG as the last command of the DDT string will cause

the DDT command to be repeated continuously until a

different command is sent to break the loop.

The TRG command must be the last command of the

DDT string, otherwise the input buffer will overflow,

causing an error.

Syntax:

*TRG

Parameters:

None

Example:

To cause the command in the define device trigger

buffer to be executed:

*TRG

The response from the FM5004 will be exactly the

same as if the command contained in the define device

trigger buffer were executed directly.

104 |

Remote Operation

Archived 10/19/10