Filter size, Log scale output – ETS-Lindgren FM5004 Field Monitor (Archived) User Manual

Page 45

background image

FILTER

SIZE

Calls up the numeric entry screen to select the number of samples to use for

each analog output update. The saved sample represents a sliding window and

serves as a smoothing function for the analog output signal. The analog output

level is updated every sample according to the samples per second setting in the

Main Menu.

If the filter size is greater than one, the sample window contains the

current sample plus the most recent samples.

If the filter size is set to one, only the current sample is used by the

analog output system.

The minimum/maximum/average setting affects how the data is processed. See

Min/Max/Avg on page 44.

LOG

SCALE

OUTPUT

Selects either linear or log scale for the analog output voltage.

When selected, the output will track a logarithmic curve with a zero

field level being zero volts out and a full scale field level being five volts

out.

When deselected, the endpoints will be the same but the analog

voltage out to field level ratio will be linear.