7.0 Remote Operation

The FM5004 Field Monitor may be operated remotely by connecting the

IEEE-488 parallel port or the RS-232 serial port to a remote device, such as an

IEEE-488 bus or a personal computer.

The IEEE-488 port is IEEE-488.2 compatible.

Command Set Format

See Remote Commands on page 60 for a detailed description of

remote commands.

Each command code is composed of two or three alpha characters and an

optional numeric parameter; commands can be entered in upper or lower case.

If a command is sent without the optional numeric parameter, the instrument

assumes the command is a status request. When the command is a status

request, the instrument returns a string consisting of the command followed by

the current setup for that command.

All commands are terminated by a line feed . To include several commands

in a single command string, separate each command by a semicolon (delimiter).

Do not include spaces between a command and the associated numeric

parameter. Data returned from the instrument will be terminated with a

.

IEEE-488

IEEE-488

R

EQUIREMENTS

For IEEE-488 operation, the device address is set using the menu system. At the

factory the device address is set to 4. If your IEEE-488 device uses address 4,

reset the switch to a vacant address.

Remote Operation

| 57