Alarm lower enable – ETS-Lindgren FM5004 Field Monitor (Archived) User Manual

Page 66

background image

A

LARM

L

OWER

E

NABLE

Description:

Enables the line 1, line 2, 4-probe/3-axis, battery, or

temperature lower limit alarms. Each of these alarms

must be enabled and disabled separately.

In the case of the low battery alarm, the level is not

adjustable. See the Battery on page 30 for a

description of the low battery alarm operation. See

Alarm Lower Value on page 67 to set the alarm lower

values.

Syntax:

ALOE,m,x

Parameters:

Alarm Type (m):

• 0 = low temperature alarm enable

• 1 = line 1 lower alarm enable

• 2 = line 2 lower alarm enable

• 4 = 4-probe/3-axis lower alarm enable

• 5 = low battery alarm enable

Numeric value (x):

• 0 = lower alarm disabled

• 1 = lower alarm enabled

Example:

• To enable the line 1 lower alarm:

ALOE,1,1

• To verify the line 1 lower alarm status:

ALOE,1?

The FM5004 will return:

ALOE,1,1

66 |

Remote Operation

Archived 10/19/10