Change range – ETS-Lindgren FM5004 Field Monitor (Archived) User Manual

Page 74

background image

C

HANGE

R

ANGE

Description:

Change the range of a probe.

Syntax:

CHRA,c,r

Parameters:

Channel (c):

• 1–4

Range (r):

• 1–4 (or highest range) or A (autorange)

Example:

• If command is entered with no parameters, all probe

ranges will be returned.

• If no probe is on that channel, range = 0

• If command is entered with one parameter, the

current range of that probe on that channel is

returned CHRA?. The FM5004 will return:

CHRA,1,A1
Probe on channel 1 is in autorange on range 1

CHRA,2,1
Probe on channel 2 is on range 1

CHRA,3,0
Probe on channel 3 is not on or not ready

CHRA,4,A3
Probe on channel 4 is in autorange on range 3

• To verify the range of the probe on channel 1:

CHRA,1?

The FM5004 will return:

CHRA,1,A1

• If command is entered with two parameters the

probe on that channel will be set to that range.

CHRA,1,2
Set probe on channel 1 to range 2

CHRA,2,A
Set probe on channel 2 to autorange