Calibration and setup, cont – Cardinal Industries 225 WEIGHT INDICATOR User Manual
Page 71

CALIBRATION AND SETUP, CONT.
Dual-Point with Zero (First Zero) Calibration
on the
U.
CAL 1= (FIRST CALIBRATION WEIGHT)
With the SCALE 1 CALIBRATION MENU displayed, press the 1 and 0 keys and then the
ENTER key. The display will change to show the settings for capacity, interval, decimal
point position and
CAL 1=0.0. This is the first of two calibration weights. It could be
ZERO (no load) or the TEST load (weight).
1. If the first calibration weight is to be ZERO (no load), press the ENTER key. The display
will flash
CALIBRATING... for a few seconds and then change to show CAL 2=.
2. If the first calibration weight is to be the TEST load (weight), place the weights on the
scale.
3. Using the numeric keys, enter the value of the TEST load (weight) and then press the
ENTER key.
4. The display will flash
CALIBRATING... for a few seconds and then change to
show
CAL 2=.
CAL 2= (SECOND CALIBRATION WEIGHT)
The display will show
CAL 2=XXXXXXX. This is the second of two calibration weights.
It could be ZERO (no load) or the TEST load (weight).
1. If the second calibration weight is to be ZERO (no load), press the 0 key and then the
ENTER key. The display will flash
CALIBRATING... for a few seconds and then
change to show
CAL 3=.
2. If the second calibration weight is to be the TEST load (weight), place the weights
scale.
3. Using the numeric keys, enter the value of the test load (weight) and then press the
ENTER key.
4. The display will flash
CALIBRATING... for a few seconds and then change to
show
CAL 3=.
CAL 3= (LAST CALIBRATION WEIGHT)
The display will show
CAL 3=. This weight is not used. Press the ENTER key to skip
CAL 3=. The display will change to show the SCALE 1 CALIBRATION MEN
8200-M538-O1
y 225 Installation
Page 67