Calibration and setup, cont – Cardinal Industries 225 WEIGHT INDICATOR User Manual
Page 41
![background image](/manuals/60136/41/background.png)
CALIBRATION AND SETUP, CONT.
4. OIML=XXX (PROTECTIVE INTERFACE)
NOTE! If the 225 is used in a country where the application requires a protective
interface, the
USA= parameter must be set for international use (USA=NO) to allow
the
4. OIML=XXX setup parameter selection.
With SETUP MENU #1 displayed the current setting for the
OIML= parameter will be
shown. Note that XXX is the current value. If the setting displayed is acceptable, procee
to the next setup param
d
eter.
NOTE
ES is selected, serial port COM1 will be dedicated to outputting scale gross
eight whenever the application is executing.
5.
the current value. If the setting displayed is acceptable, proceed to
e next setup parameter.
r 24
1
24 h
(8:3
6. CL
With SE
rrent settin
will be
shown. Note that XXX is the current value. If the s
ceptable, proceed
to the n
Otherw
, the ENTER key, YES
NO (on the soft keys) and then the
ENTER
setting and return to SETUP MENU #1.
CLR TARE=NO
below zero
Stored Tare is not cleared when
Net weight goes below zero
Otherwise, press the 4 key, the ENTER key, YES or NO (on the soft keys) and then the
ENTER key to save the new setting and return to SETUP MENU #1.
OIML=YES OIML=NO
Protective Interface Enabled
! When Y
Protective Interface Disabled
w
TIME=XX (CLOCK TIME FORMAT)
With SETUP MENU #1 displayed the current setting for the
TIME= parameter will be
shown. Note that XX is
th
Otherwise, press the 5 key, the ENTER key and then using the numeric keys, enter 12 o
nd then press the ENTER key to save the new setting and return to the SETUP MENU #1.
a
TIME=12 TIME=24
2 hour clock (AM/PM)
(8:30 PM will print as 08:30 PM)
our clock
0 PM will print as 20:30)
R TARE=XXX (CLEAR TARE)
TUP MENU #1 displayed the cu
g for the
CLR TARE= parameter
etting displayed is ac
ext setup parameter.
ise, press the 6 key
or
key to save the new
CLR TARE=YES
ears Stored Tare
Automatically cl
when Net weight goes
The following is a typical example of the Clear Tare feature in use.
1. Pla
then press the TARE key.
2. Load container with item to be weighed and perform normal weighing operation.
3. Re
4. Scale weight returns to below zero (weight of container) and is then reset to zero.
5. Operator is required to repeat step 1 before next weighing operation.
ce container on scale,
move load (item AND container) from scale.
8200-M538-O1
y 225 Installation
Page 37