Calibration and setup, cont – Cardinal Industries 225 WEIGHT INDICATOR User Manual
Page 62
CALIBRATION AND SETUP, CONT.
9. WT INTERVAL=X (WEIGHT INTERVAL “SINGLE OR DUAL”), CONT.
all to provide adequate signal
strength for stable operation of the indicator. Most scale installations will allow the
The inherent sensitivity and capabilities of the 225 will allow you to select combinations that
are beyond the practical limits of today’s load cells for dependable, stable performance. The
ap
wer range, while the higher range will
operat
With SCALE 1 SETUP MENU #1 displayed, press NEXT
to p
CAUTION! When using the dual interval feature of the 225, do not attempt to use
a division value in the low range that is too sm
smaller division to be 1/2 of the larger division value when the dual interval feature
is used.
pearance of such a situation will be instability in the lo
e as it should.
(Navigation Keys
Down Arrow)
roceed to SCALE 1 SETUP MENU #2.
SCALE 1 SETUP MENU #2
1. FILTER=X
6. CAP=XXXXXXX
2. FMAX=XXX
3. FBRK=XXX
4. INTERVAL=XX
5. DPP=X
Enter Selection: 0
`
NEXT
^
EXIT
NOTE! Although shown on the scale setup menu, the setup parameters
FMAX=
and
FBRK= can only be selected when CUSTOM FILTERING (FILTER=3)
has been selected.
r 3.
)
amp
2
MODERATE FILTERING (Sample Rate = 1)
3
CUSTOM FILTERING (additional parameters available)
1. FILTER=X (DIGITAL FILTERING)
With SCALE 1 SETUP MENU #2 displayed the current setting for the
FILTER=
parameter will be shown. Note that X is the current value for the digital filtering. If the
setting displayed is acceptable, proceed to the next setup parameter.
Otherwise, press the 1 key, the ENTER key and then using the numeric keys, enter the
value for the digital filtering and then press the ENTER key to save the new setting and
return to SCALE 1 SETUP MENU #2. Allowable values are 0, 1, 2, o
0
DISABLED (NO Filtering
1
MINIMAL FILTERING (S
le Rate = 2)