Tro t. ubleshooting, con – Cardinal Industries 225 WEIGHT INDICATOR User Manual
Page 108
![background image](/manuals/60136/108/background.png)
TRO
T.
UBLESHOOTING, CON
COLD RAM
The eXternal ram has failed to retain memory.
U
CORRECTIVE ACTION
U
: Consult your scale service provider.
CAL LOST CONFIG REQUIRED SCALE X
ates improper stored calibration data, calibration is
number (1, 2 or 3).
U
EEPROM checksum failure. Indic
necessary. Note that X is the scale
CORRECTIVE ACTION
U
OR
: Recalibrate with calibrated test weight.
ERR
ttempted:
A.
B. TARE key pressed to enter a push button tare value of a negative value.
pressed to enter a tare weight value that exceeds the scale capacity.
tare weight value that is inconsistent with the scale
ale divisions of 5).
E.
ge to kilograms when the kilogram tare weight value exceeds
U
CO
An invalid keypad entry was a
PRINT key pressed with a negative weight.
C. ENTER key
D. ENTER key pressed to enter a
division value (i.e. attempt to enter a tare of 123 with sc
ZERO key pressed when the gross weight is outside the scale zero weight range.
F. lb/kg key pressed to chan
4 digits in length.
RRECTIVE ACTION
U
error display is applicable
e appropriate corrective action.
¤
ducing weight.
: Determine which of the reasons for the
and take th
A2D ERROR
¤ ¤ ¤ ¤ ¤ ¤
The Analog to Digital converter chip has stopped pro
U
CORRECTIVE ACTION
U
: Consult your scale service provider.
DRDY ERROR
¤ ¤ ¤ ¤ ¤ ¤ ¤
The Analog to Digital converter chip is not responding.
U
CORRECTIVE ACTION
U
: Consult your scale service provider.
LE FULL
I
attempting to add an ID when the ID file is full.
F
The operator is
U
CORRECTIVE ACTION
U
: Determine the reason for the error display and take the
appropriate corrective action.