Calibration and setup, cont – Cardinal Industries 225 WEIGHT INDICATOR User Manual
Page 61
![background image](/manuals/60136/61/background.png)
CALIBRATION AND SETUP, CONT.
6. SAMPLES=XXX (SAMPLE RATE)
With SCALE 1 SETUP MENU #1 displayed the current setting for the
SAMPLES=
parameter will be shown. Note that XXX is the current value. If the setting displayed is
acceptable, proceed to the next setup parameter.
Otherwise, press the 6 key, the ENTER key and then using the numeric keys, enter the
value for the sample rate in samples per second and then press the ENTER key to sa
the new setting and return to SCALE 1 SETUP MENU #1. Allowable values fo
ve
r the
sample rate are: 1 through 100.
7. STABLE=XX (STABLE “MOTION RANGE”)
W
ETUP ME
the curre t setting for the
STABLE=
parameter will be shown.
e curre t value for the range of motion
detection. If the setting displayed is acceptable, proceed to the next setup parameter.
the
he new setting and return to
CALE 1 SETUP MENU #1. Allowable values are: 0 through 99 divisions.
8.
ber of consecutive stable
eight readings before indicating stable weight. If the setting displayed is acceptable,
R
turn to SCALE 1 SETUP MENU #1. Allowable values are: 1 through 255.
9.
R DUAL”)
ith SCALE 1 SETUP MENU #1 displayed the current setting for the
W
arameter will be sho
the current value. If the
sett
table, proceed to the next setup parameter.
using the numeric keys, enter the
nd
WT INTERVAL=1
WT INTERVAL=2
ith SCALE 1 S
NU #1 displayed
n
Note that XXX is th
n
Otherwise, press the 7 key, the ENTER key and then using the numeric keys, enter
value for the new range (the number of divisions of change permitted before indicating
unstable) and then press the ENTER key to save t
S
STBL CNT=XXX (STABLE COUNT “CONSECUTIVE STABLE READINGS”)
With SCALE 1 SETUP MENU #1 displayed the current setting for the
STBL CNT=
parameter will be shown. Note that XXX is the current num
w
proceed to the next setup parameter.
Otherwise, press the 8 key, the ENTER key and then using the numeric keys, enter the
value for the stable count and then press the ENTE key to save the new setting and
re
WT INTERVAL=X (WEIGHT INTERVAL “SINGLE O
W
T INTERVAL= p
wn. Note that X is
ing displayed is accep
Otherwise, press the 9 key, the ENTER key and then
value for the weight interval and then press the ENTER key to save the new setting a
return to SCALE 1 SETUP MENU #1. Allowable values are: 1 or 2.
Single Interval (One weight range)
Dual Interval (Two weight ranges)
8200-M538-O1
y 225 Installation
Page 57