Troubleshooting, Error codes – Cardinal Industries 225 WEIGHT INDICATOR User Manual
Page 107
TROUBLESHOOTING
Error Codes
The Model 225 indicator is equipped with software that indicates when an error in the
operation takes place. The following lists the error codes displayed by the 225 along with their
, please refer to this list for the cause.
¤ ¤ ¤ ¤
d range of the indicator.
meaning. Should you encounter an error code
ANALOG HIGH
¤ ¤ ¤
1. The load cell input is above the allowe
U
CORRECTIVE ACTION
U
: Check for imprope
output of 1 to 40mV.
Load cell or circuit failure.
r load cell wiring, excessive load, and for
2.
U
CORRECTIVE ACTION
U
: Consult your scale
service provider.
1. The load cell input is below the allowed range of the ind
N
ANALOG LOW
¤ ¤ ¤ ¤ ¤ ¤ ¤
icator.
U
CORRECTIVE ACTIO
U
: Check for improper load cell wiring and for output of 1 to 40mV.
2. Load cell or circuit failure.
U
CORRECTIVE ACTION
U
: Consult your scale service provider.
lus 9
U
CORRECTIVE ACTION
OVERCAP
¤ ¤ ¤ ¤ ¤ ¤ ¤
The load on the scale exceeds the scale capacity (105% of capacity if
USA=YES or p
divisions if
USA=NO). May indicate miscalibration.
U
: Remove the over capacity load from the scale platform. Consult
your scale service provider regarding recalibration.
CHECK MATL
In a Digital Fill Control or Batcher operation configured for decumulative operation
(
ACCUMLATE=YES) there is not enough material to start the operation.
U
CORRECTIVE ACTION
U
: Determine the reason for the error display and take the
appropriate corrective action.
CHECK ZERO
In a Digital Fill Control or Batcher operation with
DUMP GATE=YES and AUTO
TARE=NO the weight is above the zero tolerance value.
U
CORRECTIVE ACTION
U
: Determine the reason for the error display and take the
appropriate corrective action.