beautypg.com

B&G Deckman User Manual

Page 184

background image

User variables

12.25

Note that from inside Deckman you are only able to calibrate either one
of the slope and intercept (see Figure 12.1 Figure 12.1), to change both
you must used the advanced facilities outlined in this chapter.

A file might look like

Cal_Cmp

K

82

5

1

0

=====================
Parameters:

1: name (appears on Deckman)
2:

code

(K

for

single

value

calibration)
3:

variate

number

in

main

list

(j_varsXX.d)
4: width of display field in chars
5: number of decimal places
6: cal param id (0 intercept, 1 slope,
2 inverted slope, 3 set value)

The result in Deckman
Inside Deckman, hit gmenu>user vars control will bring up a dialog
like those shown (Figure 12.2 and Figure 12.3). The button at the top of
the window allows you to choose either calibration or damping.

The example on the right is the damping control, and operates in
seconds (rather than being multiplied by 5Hz in the file).