Scaling analog inputs for 0-2 ma and ±2 ma, Programming analog outputs – SATEC PM174 Manual User Manual
Page 76

Chapter 4 PAS Application
General Meter Setup
76
Series PM174 Powermeters
zero. The device does not allow you to access this setting.
Whenever the direction of the input current is changed to
negative, the device automatically uses your full engineering
scale settings for +1 mA with a negative sign.
Scaling Analog Inputs for 0-2 mA and ±2 mA
The input scales for 0-1 mA and ±1 mA analog inputs are
always programmed for 0 mA and +1 mA regardless of the
desired input range. If you want to use the entire input range of
2 mA or ±2 mA, set the analog input scales in your device as
follows:
0-2 mA: set the 1 mA scale to 1/2 of the required full scale
output for uni-directional parameters, and set the 0 mA
scale to the negative full scale and the 1 mA scale to zero
for bi-directional parameters.
±2 mA: set the 1 mA scale to 1/2 of the required full-scale
output for both uni-directional and bi-directional
parameters.
For example, to convert voltage readings from the analog
transducer that transmits them in the range of 0 to 2 mA to the
range 0 to 120V, set the full range for the +1 mA analog input
to 60V; then the 2 mA reading is scaled to 120V.
Programming Analog Outputs
The meter can be ordered with two optional analog outputs
with options for 0-1mA, ±1mA, 0-20mA or 4-20mA output
currents.
The 0-1mA and ±1mA current outputs provide a 100%
overload, and actually output currents up to 2 mA and ±2mA
whenever the output value exceeds the engineering scale set
for the 1 mA or ±1mA.
To configure the Analog Outputs in your device, select
General Setup from the Meter Setup menu, and then click on
the Analog Outputs tab. If you are programming your device
online, analog outputs are designated as not available if they
are not present in the device.
The available analog output options are described in the
following table.
Option Range
Description
AO type
0-1mA
±1mA
0-20mA
4-20mA
The analog output type. When connected to the
meter, shows the actual AO type read from the
device. When working off-line, select the analog
output option corresponding to your meter.
Output
parameter
See Appendix B
Selects the measured parameter to be transmitted
through the analog output channel.
Zero scale
Defines the low engineering scale (in primary units)
for the analog output corresponding to a lowest
(zero) output current (0 or 4 mA)
Full scale
Defines the high engineering scale (in primary units)
for the analog output corresponding to a highest
output current (1 or 20 mA)