21

Option Slot 3 - SSR Driver Output Module

If option slot 3 is fitted with an

SSR

driver output module, make connections as illustrated. The solid-

state relay driver is a 0-10V DC signal; load impedance must be no less than 500 ohms. SSR driver
outputs are not isolated from the signal input or other SSR driver outputs.

Figure 24. Option Slot 3 - SSR Driver Module Connections

Option Slot 3 - Linear Voltage or mADC Output module

If option slot 3 is fitted with a DC linear output module, make connections as illustrated.

Figure 25. Option Slot 3 - Linear Voltage & mADC Module Connections

Option Slot 3 - Transmitter Power Supply Module

If option slot 3 is fitted with a transmitter power supply module, make connections as illustrated. The
output is an unregulated 24V DC, 22mA supply.

Figure 26. Option Slot 3 - Transmitter Power Supply Module Connections

WARNING:

THIS MODULE MUST NOT BE FITTED INTO OPTION SLOT 2.