beautypg.com

Dakota Ultrasonics MINI--MAX User Manual

Page 90

background image

Dakota Ultrasonics

84

2) Use the UP and DOWN arrow keys to scroll through the sub menu

items until OPEN is highlighted.

3) Press the ENTER key to display the Open Group List Box.

4) Use the UP and DOWN arrow keys to scroll through the list menu items

until the group name containing the probe zero data is highlighted.

5) Press the ENTER key to load the zero data group into memory, followed

by pressing the OK key to confirm loading the group into memory.

Note: The user will always use this group for the zero verification

procedure, as it contains the original settings and measurement data of the

probe zero calibration standards.

6) Apply a drop of couplant to the cal standard or transducer, and attach it

to one end of the cal standard. Rotate the transducer clockwise and

counter clockwise applying a small amount of pressure to eliminate any

excess couplant between the transducer and cal standard. Be sure to

always place the transducer in the same location. This will help to

eliminate any potential measurement errors caused by changing the

sound path.