Dakota Ultrasonics MINI--MAX User Manual
Page 124
![background image](/manuals/557266/124/background.png)
Dakota Ultrasonics
118
17) Measure and record the Ultrasonic Length at Load 3 by pressing the
ENTER key. The Digits Edit Box will be displayed.
Note: Load 3 should be stored in column D.
18) Enter the Known Load value by pressing the UP and DOWN arrow
keys to scroll the highlighted value.
19) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
20) Repeat steps 18 & 19 until the Known Load value is correctly
displayed.
21) Press the OK key to store the Known Load and return to the
measurement screen, or ESC to cancel entering the Known Load.
22) Repeat Steps 1 – 21 for sample bolts 2 & 3.
Now that we have measured and documented three sample bolts, we’re now
ready to run the vector or regression calculation to determine a load factor and
offset (regression only).
Running a Regression or Vector
1) Press the MENU key multiple times to tab right and the ESC key
multiple times to tab left until the CAL menu is highlighted and
displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until LOAD CAL CALC is highlighted.