Starting delay
The starting DELAY, or starting length, is the value displayed on the bottom
lower left side of the display in both RF and RECT (Rectified) views. The DELAY
will also be represented in the DIGITS (Scan Bar) view, as the length. The
starting length is the minimum length that can be viewed on the display.
Note: Once the range is set, it will remain the same for all the views
respectively.
The procedures to adjust the Delay are outlined below:
Adjusting the Delay using the Hot Menus
1) Press the MEAS key once to activate measure menu items. Press the
MEAS key multiple times to move right and the ESC key multiple times
to move left, until the DELAY cell is highlighted.
2) Press the UP, DOWN, LEFT, and RIGHT arrow keys to scroll the
highlighted value.
3) Alternatively, press the ENTER key to display the Digits Edit Box.
4) Press the UP and DOWN arrow keys to scroll the highlighted value.
5) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
6) Repeat steps 4 & 5 until the DELAY value is correctly displayed.