Dakota Ultrasonics MINI--MAX User Manual
Page 81
![background image](/manuals/557266/81/background.png)
MINI-MAX Ultrasonic Bolt Tension Monitor
75
most significant – 6” bar. If the glass block is being used, it will have only
one velocity specified for the block.
19) Press the ENTER key to display the Digits Edit Box.
20) Press the UP and DOWN arrow keys to scroll the highlighted value.
21) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
22) Repeat steps 20 & 21 until the VELOCITY value is correctly displayed.
23) Press the OK key to return to the Two Point Zero list, or ESC to cancel
entering the VELOCITY.
24) Press the UP and DOWN arrow keys to scroll through the Two Point
Zero list until TEMP COEF is highlighted.
Note: This should be set to 55 (99 metric) for the 3 & 6” bars supplied by
Dakota Ultrasonics. If the glass block is used, it should be set to 0. The
reason for this is because glass has a very low temperature coefficient.
25) Press the ENTER key to display the Digits Edit Box.
26) Press the UP and DOWN arrow keys to scroll the highlighted value.
27) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
28) Repeat steps 26 & 27 until the TEMP COEF value is correctly
displayed.
29) Press the OK key to return to the Two Point Zero list, or ESC to cancel
entering the TEMP COEF.
30) Press the ESC key once to leave the Two Point Zero list.
31) Press the MENU key multiple times to tab right and the ESC key
multiple times to tab left until the MATL menu is highlighted and
displaying the submenu items.