Dakota Ultrasonics MINI--MAX User Manual
Page 84
Dakota Ultrasonics
78
4) Press the MENU key multiple times to tab right and the ESC key
multiple times to tab left until the AUTO menu is highlighted and
displaying the submenu items.
5) Use the UP and DOWN arrow keys to scroll through the sub menu
items until APROX LEN is highlighted.
Note: This is the approximate physical length of the cal standard currently
being measured. This tells the MINI-MAX where to turn on the receiver and
look for an echo. The area scanned will be +/- 5% from the approximate
length. For this example, 3” & 6” cal bars are used. Start with the 3” cal bar
first. Note: If the glass block is being used as the standard, use the 1” side.
6) Press the ENTER key to display the Digits Edit Box.
7) Press the UP and DOWN arrow keys to scroll the highlighted value.
8) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
9) Repeat steps 7 & 8 until the APPROX LEN value is correctly displayed.
10) Press the OK key to set the Approximate Length and return to the
menu screen, or ESC to cancel entering the Approximate Length.
Note: Alternatively, the Approximate Length can be entered from the
main measurement screen using the ALEN measure menu item. This can
be accessed by pressing the MEAS button on the keypad multiple times
until the ALEN field is highlighted.
11) Use the UP and DOWN arrow keys to scroll through the sub menu
items until AUTO SET is highlighted.