MINI-MAX Ultrasonic Bolt Tension Monitor
77
Setting up the MINI-MAX for Measurement
Note: Check to see if the system was previously calibrated. You should
NOT change the calibration unless the transducer, transducer cable, or
instrument has been changed. In any event, the probe zero calibration
should always be checked prior to changing it and only changed if it is
required, as you can introduce error by uncertainty during the zero
procedure. If the MINI-MAX has been previously calibrated, open the
calibration data saved in the MINI-MAX and check the probe zero by simply
measuring the bars in an empty location (columns B – ZZ) in the existing
probe zero data group. If the zero is slightly off, minor adjustments to the
temperature, located in the Main Measure Menu, should be all that’s
necessary to bring the error between the two bars within a tolerance of
.0005”.
1) With the menu items still active from the previous section, press the
MENU key multiple times to tab right and the ESC key multiple times to
tab left until the GEOM menu is highlighted and displaying the submenu
items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until QUANTITY is highlighted.
3) Use the LEFT and RIGHT arrow keys to scroll the quantity options until
the ELONG option is displayed.