MINI-MAX Ultrasonic Bolt Tension Monitor

119

3) Press the ENTER key to run the calculation and display the confirmation

screen.

Note: The regression or vector coefficient will be displayed. The closer

this number is to 1, the better the fit, or least amount of error.

4) Press the OK key to accept the calculation, or the ESC key to cancel

and return to the CAL menu.

5) Press the MEAS key to return to the measurement screen.

Note: The new Load Factor and Offset values can be reviewed and

edited in the GEOM tabbed menu item.

6) Calibration is completed.