Dakota Ultrasonics MINI--MAX User Manual
Page 125
MINI-MAX Ultrasonic Bolt Tension Monitor
119
3) Press the ENTER key to run the calculation and display the confirmation
screen.
Note: The regression or vector coefficient will be displayed. The closer
this number is to 1, the better the fit, or least amount of error.
4) Press the OK key to accept the calculation, or the ESC key to cancel
and return to the CAL menu.
5) Press the MEAS key to return to the measurement screen.
Note: The new Load Factor and Offset values can be reviewed and
edited in the GEOM tabbed menu item.
6) Calibration is completed.