MINI-MAX Ultrasonic Bolt Tension Monitor

73

Measuring Calibration Standards – For the First Time

1) With the menu items still active from the previous section, press the

MENU key multiple times to tab right and the ESC key multiple times to

tab left until the CAL menu is highlighted and displaying the submenu

items.

2) Use the UP and DOWN arrow keys to scroll through the sub menu items

until ZERO MODE is highlighted.

3) Use the LEFT and RIGHT arrow keys to scroll the zero mode options

until the TWO POINT option is displayed.

Note: The two-point option allows the user to use two separate points for

the probe zero calibration. Note: The one point option can optionally be

selected if only one cal bar will be used. Although the procedures below

are described with the two-point option in mind, they will also work in the

case of a one-point calibration.

4) Use the UP and DOWN arrow keys to scroll through the sub menu items

until MEASURE ZERO is highlighted.

5) Press the ENTER key to display the Two Point Zero list box.

6) Press the UP and DOWN arrow keys to scroll through the Two Point

Zero list until PHY LEN 1 is highlighted.