MINI-MAX Ultrasonic Bolt Tension Monitor

151

may be convenient for those users who want to have a quick verification without
having to constantly watch the display for a specific value to appear. The alarm
mode uses red and green LED’s to let the user know when the fastener is inside
the tolerance limits. The limits are also displayed on the LCD in terms of a bar
graph presentation. There are two limits, or alarm values, that can be setup in
the MINI-MAXALARM LO LIMIT and ALARM HI LIMIT. However, the user
may choose to activate and utilize only one of the limit values, depending on their
specific application requirements. The procedures to use the ALARM MODE
feature are outlined below:

UTIL MENU – ALARM MODE

Toggle Alarm Status (on/off)

1) Press the MENU key once to activate the menu items tab. Press the

MENU key multiple times to tab right, and the ESC key multiple times to

tab left, until the UTIL menu is highlighted and displaying the submenu

items.

2)

Use the UP and DOWN arrow keys to scroll through the sub menu items

until ALARM STATUS is highlighted.