10) Repeat steps 8 & 9 until the PHYSICAL LEN value is correctly

displayed.

11) Press the OK key to return to the One Point Zero List, or ESC to

cancel entering the PHYSICAL LEN value.

12) Use the UP and DOWN arrow keys to scroll through the One Point

Zero List Items until VELOCITY is highlighted.

13) Press the ENTER key to display the Digits Edit Box.

Note: Enter the velocity value of the bolt that was documented in the

previous section. Reminder: The value was written down and scribed on

the side of the bolt. It is the velocity used for the probe zero bolt.

14) Press the UP and DOWN arrow keys to scroll the highlighted value.

15) Press the LEFT and RIGHT arrow keys to scroll the digit locations.

16) Repeat steps 14 & 15 until the VELOCITY value is correctly displayed.

17) Press the OK key to return to the One Point Zero List, or ESC to

cancel entering the VELOCITY value.

18) Use the UP and DOWN arrow keys to scroll through the One Point

Zero List Items until TEMP COEF is highlighted.