Dakota Ultrasonics MINI--MAX User Manual
Page 80
![background image](/manuals/557266/80/background.png)
Dakota Ultrasonics
74
Note: PHY LEN 1 is the actual physical length of the first calibration
standard. If the glass block is being used, the 1” side should be used for
PHY LEN 1.
7) Press the ENTER key to display the Digits Edit Box.
8) Press the UP and DOWN arrow keys to scroll the highlighted value.
9) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
10) Repeat steps 8 & 9 until the PHY LEN 1 value is correctly displayed.
11) Press the OK key to return to the Two Point Zero list, or ESC to cancel
entering the PHY LEN 1.
12) Press the UP and DOWN arrow keys to scroll through the Two Point
Zero list until PHY LEN 2 is highlighted.
Note: PHY LEN 2 is the actual physical length of the second calibration
standard. If the glass block is being used, the 3” side should be used for
PHY LEN 2.
13) Press the ENTER key to display the Digits Edit Box.
14) Press the UP and DOWN arrow keys to scroll the highlighted value.
15) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
16) Repeat steps 14 & 15 until the PHY LEN 2 value is correctly displayed.
17) Press the OK key to return to the Two Point Zero list, or ESC to cancel
entering the PHY LEN 2.
18) Press the UP and DOWN arrow keys to scroll through the Two Point
Zero list until VELOCITY is highlighted.
Note: This refers the velocity of the probe zero cal bars. Since there is a
velocity for each cal bar, use the velocity of the longest cal bar as it will be