Dakota Ultrasonics MINI--MAX User Manual
Page 62
![background image](/manuals/557266/62/background.png)
Dakota Ultrasonics
56
used to obtain the probe zero data for the probe zero calibration bolt. In this
case, calculating the probe zero value forces the new hardware to match with the
original. If the hardware has not changed, the user should not use this feature. If
the hardware hasn’t changed only the temperature value should be adjusted to
match the existing zero values.
The next two sections outline first, the procedure for calculating a probe zero
because the hardware has changed, and second, by adjusting the temperature
value because the hardware is the same and the values do not match.
Performing a Probe Zero Calibration
Note: User has selected this section because the MINI-MAX, transducer,
or transducer cable has changed.
1) Press the MENU key once to activate the menu items tab. Press the
MENU key multiple times to tab right, and the ESC key multiple times to
tab left, until the CAL menu is highlighted and displaying the submenu
items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu
items until ZERO OPTION is highlighted.
3) Press the LEFT and RIGHT arrow keys to scroll the zero mode options
until ONE POINT is displayed.
4) Use the UP and DOWN arrow keys to scroll through the sub menu items
until MEASURE ZERO is highlighted.