Dakota Ultrasonics MINI--MAX User Manual
Page 86
![background image](/manuals/557266/86/background.png)
Dakota Ultrasonics
80
17) Press the UP and DOWN arrow keys to scroll through the Two Point
Zero list until ULTRASONIC 1 is highlighted – 3” bar or 1” side of glass
block.
18) Press the ENTER key to store the measured value of the 3” bar or 1”
side of glass block.
19) Apply a drop of couplant to the bolt or transducer, and attach it to one
end of the 6” calibration bar (or 3” side of glass block). Rotate the
transducer clockwise and counter clockwise applying a small amount of
pressure to eliminate any excess couplant between the transducer and
cal bar surface. Remove and replace the transducer a couple of times
in the exact same location on the cal bar, rotating it each time to obtain
the minimum value measured. Be sure to always place the transducer
in the same location. This will help to eliminate any potential
measurement errors caused by changing the sound path.
20) Press the UP and DOWN arrow keys to scroll through the Two Point
Zero list until ULTRASONIC 2 is highlighted – 6” bar or 3” side of glass
block.
21) Press the ENTER key to store the measured value of the 6” bar or 3”
side of glass block.
22) Press the UP and DOWN arrow keys to scroll through the Two Point
Zero list until CALC ZERO is highlighted.
23) Press the ENTER key to calculate the zero.
24) Press the OK key to confirm “overwrite settings”.
25) Press the ENTER key once again to open the TWO POINT ZERO list
and evaluate the error.