MINI-MAX Ultrasonic Bolt Tension Monitor
5
Selecting the Units
Note: The Quick Start setup, selected in the previous section,
automatically defaults to english units – inches. Follow the procedure
below to change the units to metric, if needed.
1) Press the MENU key multiple times to tab right and the ESC key
multiple times to tab left until the MATL menu is highlighted and
displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until UNITS is highlighted.
3) Press the LEFT and RIGHT arrow keys to scroll through the unit options
IN (inches), IN HR (inches hi resolution), MM (millimeters), and MM HR
(millimeters hi resolution).
Note: The high-resolution options display an additional digit of resolution to
the measurement. This mode is typically used when measuring very short
bolts with very little elongation.
Selecting the Material Type