Dakota Ultrasonics MINI--MAX User Manual
Page 146

Dakota Ultrasonics
140
3) Alternatively, press the ENTER key to display the Digits Edit Box.
4) Press the UP and DOWN arrow keys to scroll the highlighted value.
5) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
6) Repeat steps 4 & 5 until the GATE value is correctly displayed.
7) Press the OK key to return to the measurement screen, or ESC to
cancel entering the GATE value.
The user can also access and adjust the gate from the tabbed menus. However,
this method is more tedious than making the adjustments using the Hot Menus.
The procedure using the tabbed menus is outlined below:
Adjusting the Gate using the Tabbed menus
1) Press the MENU key once to activate the menu items tab. Press the
MENU key multiple times to tab right and the ESC key multiple times to
tab left until the AUTO menu is highlighted and displaying the submenu
items.