3 manual adjustment, Internal adjustments, Manual adjustment – RADWAG PS 510.R2.CT Carat Balance User Manual
Page 63
- 63 -
•
P1.5
AUTO. CALIB. T.
– Temperature of automatic adjustment
(balances with internal adjustment)
Determination of factor which signals start of automatic internal adjustment
NONE
-
no
factor
will
cause
start
of
adjustement
(adjustment automatically turned off)
TEMP.
- adjustment triggered by change of temperature
TIME
- adjustment triggered by time interval set in P1.6 AUTO CALIB.
C.
BOTH
- adjustment triggered by time and temperature.
CAUTION
Function unavailable in non-verified balances.
•
P1.6
AUTO CALIB. C. - Time of automatic adjustment (balances
with internal adjustment system)
Determination of time interval, after which automatic adjustment process is
initiated.
CAUTION
Function unavailable in verified balances.
14.3
Manual adjustment
14.3.1. Internal adjustments
Press Cal button
or
The balance performs internal adjustment automatically. During
adjustment process do not load the weighing pan with any load. The
message is displayed <DO NOT TURN OFF CALIBRATION> in the
bottom line. Once adjustment process is completed, balance saves its
result in memory and returns to weighing mode.