RADWAG PS 510.R2.CT Carat Balance User Manual
Page 62
- 62 -
or
•
P1.2
EXT. CALIB.
– External adjustment (with an external weight)
Adjustment with external weight, which value is saved in factory menu of
balance, function unavailable in verified balances.
CAUTION
Function unavailable in verified balances.
•
P1.3
USER CALIB.
– User adjustment (with an external weight)
Adjustment with external weight at any mass in the balance range, however,
not lower than 30% of max range.
CAUTION
Function unavailable in verified balances.
•
P1.4
CALIB. TEST
– Adjustment test
This fuction enables comparing the result of internal automatic adjustment
with the value of internal weight saved in balance’s factory parameters.
The process is conducted automatically and its result is shown on the
display (if the balance is connected via RS 232 port with a computer or
printer, the adjustement result will be printed out). Press ESC to go back to
the previous menu.