Value release, Autozero function – RADWAG PS 510.R2.CT Carat Balance User Manual
Page 42

- 42 -
10.4.2. Value release
Due to various ambient conditions at workplaces, and in order to set balance to
present conditions, it is necessary to determine the value release parameter
as: FAST.+REL., FAST or RELIABLE. Depending on selected option,
weighing time is either shorter or longer.
10.4.3. Autozero function
In order to ensure balance’s precise mass indication, autozero (Auto) software
parameter has been introduced. The application of this function is automatic
control and correction of zero indication. When the function is enabled, it
compares balance indications at declared time interval e.g. 1s, on condition
that weighing pan is unloaded and display indication is close to zero. If results
vary less than declared AUTOZERO range e.g. one division, balance will zero
automatically, marker of stable measurement result–
and precise zero
marker –
will be displayed.
If AUTOZERO function is enabled, then each weighing process starts from
precise zero point. There are, however, some case when this function can be a
disturbing factor in the measuring process; for instance very slow placing of
load on the weighing pan (e.g. load pouring) in such case zero indication
correction can also correct actual indication of loaded mass.