RADWAG PS 510.R2.CT Carat Balance User Manual
Page 161
- 161 -
K0
Unlock balance keypad
OMI
Give available working modes
OMS
Set working mode
OMG
Give current working mode
BP
Activate sound signal
PC
Send all implemented commands
CAUTION:
1.
Each command must end with CR LF characters; the spaces provided in
the formats should be omitted, they are included only to improve
readability.
22.2
Responses format for commands sent from computer level
On receipt of a command, the terminal sends a response in one of the following
formats:
XX_A CR LF
command understood and in progress
XX_D CR LF
command carried out (appears only after the command XX_A)
XX_I CR LF
Command understood but not accessible at this moment
XX _ ^ CR LF
Command understood but max range is exceeded
XX _ v CR LF
Command understood but min range is exceeded
XX _ OK CR LF
Command carried out
ES_CR LF
Command not recognised
XX _ E CR LF
An error occurred on command carrying out (time limit
exceeded while waiting for stable measurement result), time
limit is balance’s characteristic parameter
XX
- stands for a name of sent command
_
- stands for spaces