- 118 -
sample first in the air (top weighing pan (4) of the density kit), and then
weighing the same sample in liquid (on the bottom weighing pan (2) of the
density kit). As the same sample is weighed in liquid, the result of density
determination is automatically indicated on balance display.
In order to determine the density you need to:
1. Instal the density determination kit,
2. Enter
3. Prepare a sample,
4. Initiate the process
5. Insert the settings for the process following the displayed messages,
6. Liquid in which the determination is carried out
7.
After selecting and entering a liquid type by pressing
program will take another step, namely setting the temperature