2 menu adjustment settings, Menu adjustment settings – RADWAG PS 510.R2.CT Carat Balance User Manual
Page 61

- 61 -
•
Adjustment triggered by elapsing time. It is possible to declare time
intervals which are criteria for balance adjustment. Accessible
settings are: adjustment after 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12 hours
since last carried out adjustment procedure,
CAUTION
Declaring time intervals for balance adjustment is available only in
non-verified balances. In case of verified balances the time interval
between the following adjustments is set for 3 hours.
14.2
Menu adjustment settings
P1 CALIBRATION
P1.1 INT.CALIB.
|
[internal adjustment]
P1.2 EXT.CALIB.
|
[external adjustment]
P1.3 USER CALIB.
|
[user adjustment]
P1.4 CALIB. TEST
|
[adjustment test]
P1.5 AUTO. CAL. T. | NONE/TIME/TEMP./BOTH
[automatic
adjustment]
P1.6 AUTO CAL. C. |
[time of automatic adjustment]
•
P1.1
INT. CALIB.
– Internal adjustment
Start of internal adjustment process. Adjustment is conducted automatically,
and no user assistance is required. If weighing pan of the balance is loaded,
balance will display a command to unload the weighing pan. If weighing pan is
loaded with a relatively small mass (up to 4% of balance max capacity) then
adjustment process will be carried out automatically, but measuring results
may differ by the amount that was present on balance weighing pan.