Before the start of weighing process or in case of drastic change of ambient

conditions at a workstation (e.g. ambient temperature change at a workstation more

than 3°C) the balance requires adjusting. The procedure of balance adjustment is

described further in this user manual.

Before the start of weighing procedure, it is recommended to load the

balance’s weighing pan a few times with mass close to balance max

capacity,

Check if unloaded balance indicates “precise zero”

and whether

measurement is stable

, if not, press

/Delete

button,

Press UNITS button, to set a measuring unit:

[g], [mg], [ct], and also if enabled in factory menu (non-verified balances)

[kg], [lb], [oz], [ozt], [dwt], [tlh], [tls], [tlt], [tlc] [mom], [gr], [ti], [N], [baht], [tola]

Place weighed object on balance’s weighing pan and read result only on

stabilisation of measurement result,

Mass indication of a load placed on balance’s weighing pan can be tarred

for multiple Times by pressing →T/Insert (pay CAUTION not to exceed

maximal capacity of a balance by applying multiple tare function).

During times between carrying out the following measurement series do not unplug

the balance from mains. It is recommended to switch off the balance’s display by

pressing ON/OFF button. On repeated pressing of the ON/OFF button the balance

is ready for operation and carrying out the following measurements.