- 22 -
•
After removing the protective elements of the weighing pan and after
unblocking the internal calibration mechanism (only R2 series balances),
plug the indicator into the port which is on the back of the housing. Next,
plug in all the additional devices. Only then can the balance be connected
to mains (the power port is located at the back of the balance’s housing).
4.3
Levelling
Before plugging to mains, level the balance by turning the
adjustable feet in a way that the air bubble of the level is in
its central ring. The balance should be placed at the
workplace firmly on all of its feet.
4.4
Cleaning
1. Disassembly the weighing pan and other portable parts (the parts differ
depending on a balance type – see Unpacking and Assembly sections).
Be careful while disassembling the balance so as not to cause any
damages to the balance mechanism.