Set the air pressure, next press ENTER for confirmation. The
computer software proceeds to the next step, i.e. entering the first
control volume (V1) for a tested pipette.
•
Enter (V1) control volume. For fixed volume pipettes (P2.13.5
VOLUME DETERMINATION
parameter set to value <1>) this is the
one and only value to be entered. The computer software proceeds to
the next step, i.e. entering the second control volume (V2) for a tested
pipette.
•
Enter (V2) control volume. The computer software proceeds to the
next step, i.e. entering the third control volume (V3) for a tested
pipette.
•
Enter (V3) control volume.
•
Upon entering the data the computer software proceeds to pipettes
calibration.
•
Follow the displayed description in order to complete the procedure.