Continuous transmission, Printouts interval for continuous transmission – RADWAG PS 510.R2.CT Carat Balance User Manual
Page 150
![background image](/manuals/466047/150/background.png)
- 150 -
20.1.2. Continuous transmission
Continuous transmission parameter enables turning on/off continuous
transmission of a measurement.
It is possible to set one of three available options:
•
Continuous transmission in calibration unit
independently from currently chosen mass unit,
measurements are transmitted in unit set on a main
measuring device (calibration unit)
•
Continuous transmission in current unit
measurements are transmitted in current unit; the unit
changes dynamically whenever the user presses Units button
in order to modify the unit on a display.
•
Continuous transmission off
Setting instruction:
Caution:
Continous transmission may be turned on/off by means of command sent
from a computer (go to COMMUNICATION PROTOCOL chapter)
20.1.3. Printouts interval for continuous transmission
Parameter
continuous transmission.
Frequency of printouts is set in seconds, wherein the accuracy is 0.1s.
The user may set any time value ranging from 1000 to 0.1 s.
The setting is valid for continuous transmission in calibration unit and in current unit
activated from the balance (see parameter P4.2.1.2) and for continuous
transmission activated by command sent from a computer (see point 22
COMMUNICATION PROTOCOL).