Balance APP
Remove the adhesive tape. Carefully remove the balance from its packaging.
Remove all the elements necessary for correct operation from the accessory
box.
APP 10.R1 and APP 10.R2 balances are equipped with 2 weighing pans which
can be used in turns. It must be remembered that balance activation is only
possible with one weighing pan assembled at a time. Attempt of activating the
balance with two weighing pans assembled simultaneously will result in display
of an “-LH-“ error. In such case one of the pans needs to removed.
Carefully take the balance out of the packaging, remove transport protective
elements and gently place the balance onto its workplace.
•
Remove the weighing pan and remove transport protective elements as
shown in the picture above (if these elements are attached).
•
After the unpacking of the balance unscrew the bolt (1) which blocks the
calibration mechanism.
•
Assemble the weighing pan (for APP 10.R, one of the pans only).
•
Make sure the weighing pan is placed properly on the shock absorbers:
the weighing pan cannot touch the housing and should be stable.
•
Place the balance at the workplace.
•
Levell the balance (see the instruction below).
•
Using the wrench (3), after adjusting the balance level (see below); tighten
the bolt (2) on the foot (1) to the balance’s footing until the point of
resistance. Repeat the activity for each foot of the balance.