Returning to the weighing mode – RADWAG PS 510.R2.CT Carat Balance User Manual
Page 135

- 135 -
In order to operate properly, you should set the threshold in grams. It is
the indicating point beyond which the function starts to register
maximal force apllied.
•
From now on the balance registers and holds every single weighment
which is above the threshold, and is higher than the result of the
previous peak hold. If the software detects a mass above the
threshold, the highest detected indication is held on the main display
and the pictogram
A user can print the result out by pressing
button.
The start of the next process of peak hold measurement is possible only after
removing the load from the weighing pan and pressing
button. This causes
returning to the main mode
pictogram
18.11.2. Returning to the weighing mode
CAUTION: current unit may be selected only if the weighing result is not latched.
In order to select the unit, Esc button is used. If any result has already been
latched than Esc button performs deletion of the last latched Max result.