RADWAG PS 510.R2.CT Carat Balance User Manual
Page 165

- 165 -
Switch on continuous transmission in current measuring unit
Format: CU1 CR LF
Accessible responses:
CU1_I CR LF
- command understood but not accessible at this moment
CU1_A CR LF
MASS FRAME
- command understood and in progress
- response is mass value in current measuring unit
Responce format
1 2 3 4
5
6
7-15 16
17 18 19 20 21
S U I stability
marker space character mass space unit
CR LF
Switch off continuous transmission in current weighing unit
Format: CU0 CR LF
Accessible responses:
CU0_I CR LF
- command understood but not accessible at this moment
CU0_A CR LF
- command understood and carried out
Set low checkweighing threshold
Format: DH_XXXXX CR LF, where: _ - space, XXXXX – mass format
Accessible responses:
DH_OK CR LF
- command understood
ES CR LF
- command not recognised (mass format incorrect)
Set high checkweighing threshold
Format: UH_XXXXX CR LF, where: _ - space, XXXXX – mass format
Accessible responses:
UH_OK CR LF
- command understood
ES CR LF
- command not recognised (mass format incorrect)
Give value of low checkweighing threshold
Format: ODH CR LF
Accessible response: DH_MASS CR LF – command carried out
Responce format
1
2
3
4-12
13
14
15
16
17
18
19
D H space
mass
space
unit
space
CR
LF
Mass
- 9 characters with right justification
Unit
- 3 characters with left justification
Give value of high checkweighing threshold
Format: OUH CR LF
Accessible response: UH_MASS CR LF – command carried out