RADWAG PS 510.R2.CT Carat Balance User Manual
Page 167
- 167 -
Value release
Format: SS CR LF
Accessible responses:
SS_OK CR LF
- command understood, execution in progress
Command’s function is similar to function of button PRINT located on an overlay,
Internal calibration
Format: IC CR LF
Accessible responses:
IC_A CR LF
IC_D CR LF
- command understood, execution in progress
- calibration completed
IC_A CR LF
IC_E CR LF
- command understood, execution in progress
- range exceeded, time limit for awaiting for a stable result
IC_I CR LF
- command understood but not accessible at this moment
Disable automatic internal calibration of the balance
Format: IC1 CR LF
Accessible responses:
IC1_I CR LF
- command understood but not accessible at this moment
IC1_E CR LF
- operation disabled, e.g. for verified balance
IC1_OK CR LF
- command carried out
Operation is disabled for verified balances.
For non-verified balances the command inhibits internal calibration until it is enabled
via IC0 command or until the balance is turned off. The command does not modifiy
setings specifying calibration start.
Enable automatic internal calibration of the balance
Format: IC0 CR LF
Accessible responses:
IC0_I CR LF
- command understood but not accessible at this moment
IC0_OK CR LF
- command carried out
Operation is disabled for verified balances.
Lock balance keypad
Format: K1 CR LF
Accessible responses:
K1_I CR LF
- command understood but not accessible at this moment
K1_OK CR LF
- command carried out
Command locks the balance keypad (proximity sensors, touch panel) until the moment
of turning the balance off or until sending K0 command.
Unlock balance keypad
Format: K0 CR LF