2 format of data setting, 3 format of data sent on pressing of print button, Format of data setting – RADWAG PS 510.R2.CT Carat Balance User Manual
Page 158: Format of data sent on pressing of print button, Computer
- 158 -
21.2
Format of data setting
The measurement result can be sent from the balance to an external device by
pressing
button on the balance or through sending a command from the
computer.
21.3
Format of data sent on pressing of PRINT button
CAUTION
Unstable measurement print-out is not available for a verified balance.
Printout format
1
2
3
4 - 12
13
14 - 16
17
18
st
ab
ilit
y
m
ar
ker
sp
ace
char
ac
ter
M
as
s
sp
ac
e
uni
t
CR
LF
stability marker - [space] – when stable
- [?] – when unstable
- [^] - if there is an error exceeding the maximal range +
- [v] - if there is an error exceeding the maximal range -
character
- [space] – for positive or [-] negative values
mass
- 9 characters – aligned to the right
unit
- 3 characters – aligned to the left.
21.3.1. Format of data sent as response for commands generated from a
computer
After entering a command balance responds to the following first:
XX_A CR LF - command understood and in progress
XX_I
CR LF - command understood but not accessible at this
moment
XX_^ CR LF - command understood but maximum range is exceeded
XX_v
CR LF - command understood but minimum range is exceeded
XX_E CR LF - an error occurred on command carrying out (time limit
exceeded while waiting for stable measurement result),
time limit is balance’s characteristic parameter
XX
- name of the command