21.2

Format of data setting

The measurement result can be sent from the balance to an external device by

pressing

button on the balance or through sending a command from the

computer.

21.3

Format of data sent on pressing of PRINT button


CAUTION

Unstable measurement print-out is not available for a verified balance.

Printout format

1

2

3

4 - 12

13

14 - 16

17

18

st

ab

ilit

y

m

ar

ker

sp

ace

char

ac

ter

M

as

s

sp

ac

e

uni

t

CR

LF

stability marker - [space] – when stable

- [?] – when unstable

- [^] - if there is an error exceeding the maximal range +

- [v] - if there is an error exceeding the maximal range -

character

- [space] – for positive or [-] negative values

mass

- 9 characters – aligned to the right

unit

- 3 characters – aligned to the left.

21.3.1. Format of data sent as response for commands generated from a
computer

After entering a command balance responds to the following first:

XX_A CR LF - command understood and in progress

XX_I

CR LF - command understood but not accessible at this

moment

XX_^ CR LF - command understood but maximum range is exceeded

XX_v

CR LF - command understood but minimum range is exceeded

XX_E CR LF - an error occurred on command carrying out (time limit

exceeded while waiting for stable measurement result),

time limit is balance’s characteristic parameter

XX

- name of the command