RADWAG PS 510.R2.CT Carat Balance User Manual
Page 26
- 26 -
4. CAUTION: do not remove the front glass wall.
5. Remove the weighing pan, the weighing pan safety rings and the bottom
metal plates, so as not to damage the balance mechanism durin the
cleaning.
Thus prepared draft shield and glass walls can be properly cleaned. All the
operations should be done carefully. Pay special CAUTION to the place
where the weighing pan was installed: dirt and other small elements might
enter the balance construction through this openiing, which might
negatively influence the balance parametres.
This manual is related to the following products: